Leak out of in-line filter assembly (nanoACQUITY BSM) - WKB58519
Article number: 58519
SYMPTOMS
- Higher-than-normal consumption of solvent.
- Pre-sensor value during Set Pressure Diagnostic is very high (over 0.5 uL/min).
- Visible leak out of filter assembly.
ENVIRONMENT
nanoACQUITY UPLC BSM
CAUSE
Loose or faulty filter assembly.
FIX or WORKAROUND
- First try tightening the parts of the filter assembly together.
- If tightening of assembly does not resolve the issue, replace the filter assembly (289002111).
- Verify that BSM passed Set Pressure Diagnostics with expected values of pre-sensor and post-sensor flow values.
ADDITIONAL INFORMATION
Because during PM, filter assembly is opened to replace the insert, with time, it becomes loose and starts leaking.