ACQUITY H-Class QSM leak test fails - WKB58471

Article number: 58471

SYMPTOMS

  • Dynamic leak test fails

ENVIRONMENT

  • MassLynx
  • Empower
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC BSM
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
  • ACQUITY ISM

CAUSE

Pump head tubing is not tight enough.

FIX or WORKAROUND

1. Ensure that each fitting is tight.

2. Tighten each pump tubing.

ADDITIONAL INFORMATION

Do not over-tighten the tubings. Waters recommends that each fitting first be screwed on finger-tight, and then tightened an additional 3/4-turn with a wrench.