"Leak detected" error with ACQUITY solvent manager - WKB56750
Article number: 56750
SYMPTOMS
- "Leak detected" error with ACQUITY solvent manager.
- Waste container is empty and there is no air trapped in waste line.
ENVIRONMENT
- Empower
- ACQUITY UPLC BSM
CAUSE
Blockage in the BSM waste drain due to buffer precipitation.
FIX or WORKAROUND
-
Remove leak sensor and pour hot water into the BSM drain to remove buffer/residue.
-
Replace leak sensor and prime BSM to confirm that the issue is resolved.
ADDITIONAL INFORMATION