ACQUITY i2V fails primary pump head dynamic leak test - WKB56656

Article number: 56656

SYMPTOMS

  • ACQUITY pump with i2V fails dynamic leak test on primary pump head

ENVIRONMENT

  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
  • ACQUITY Classic Binary Solvent Manager with i2V
  • ACQUITY Premier Binary Solvent Manager (BSM)
  • ACQUITY UPLC I-Class Binary Solvent Manager (BSM)

CAUSE

Faulty i2V cartridge or i2V actuator assembly

FIX or WORKAROUND

Replace the i2V cartridge first, if this fails to fix the leak replace the i2V actuator assembly.

ADDITIONAL INFORMATION

H-Class QSM part numbers:

  • i2V Cartridge (1pk) - 700005165 
  • i2V Actuator assembly - 700005162

I-Class BSM part numbers:

  • i2V Cartridge (2pk) - 700004139 
  • i2V Actuator assembly - 700003557 

Classic BSM part numbers:

  • i2V Cartridge (1pk) - 700005165 
  • i2V Actuator assembly - 700003557 

Premier QSM and BSM parts numbers: 

  • i2V Cartridge - 700005414
  • i2V Actuator assembly - 700005412 