No peaks in chromatogram with 717 autosampler - WKB55966

Article number: 55966

SYMPTOMS

  • No peaks appear in chromatograms
  • Seal pack calibration passes
  • Autozeroing transducer passes

ENVIRONMENT

  • 717 Autosampler

CAUSE

Loose or faulty sample syringe

FIX or WORKAROUND

  1. Tighten the sample syringe.
  2. If sample syringe is faulty, replace it.

ADDITIONAL INFORMATION

 

id55966, 717, 717HC, 717PLUS, 717UP