600E pump is not delivering solvent - WKB55864

Article number: 55864

SYMPTOMS

  • Indicator rods on the pump heads do not move

  • Fan on the back of the pump does not move when the unit is turned on

  • When a gradient is programmed on the 600 Controller, the gradient proportioning valve in the pump is heard functioning

  • Sparging works

ENVIRONMENT

  • Modular LC
  • 600E pump

CAUSE

Bad pump power fuse

FIX or WORKAROUND

Replace FUSE,6A,125V,5X20MM FAST-ACTING UL (WAT162-23), which is located on the rear panel of the Waters 600 pump controller, as shown below.

Waters 600 Controller Rear Panel.PNG