Skip to main content
Waters

ACQUITY H-Class FTN needle readiness test fails - WKB5554

Article number: 5554

SYMPTOMS

  • ACQUITY H-Class FTN needle readiness test fails
  • Pressure is unstable
  • Retention time is longer
  • Installation of new needle

ENVIRONMENT

  • ACQUITY H-Class FTN

CAUSE

Replacement of needle without replacing the needle seat.

FIX or WORKAROUND

  1. Replace the needle seat (405011492 - SEAT, VESPEL W/ANTI-ROT).
  2. Run the needle readiness test.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.