No peaks seen when injecting standards with a 2707 Autosampler - WKB54761

Article number: 54761

SYMPTOMS

  • No peaks present on standard injection
  • No errors present on 2707 console

ENVIRONMENT

  • 2707 Sample Manager

CAUSE

The sample syringe is broken.

FIX or WORKAROUND

Replace the sample syringe (700000862).

ADDITIONAL INFORMATION

 

id54761, 2707, 2707SFC