How to set the TUV lamp serial number in the ACQUITY console - WKB5440

Article number: 5440

OBJECTIVE or GOAL

Set the TUV lamp serial number in the ACQUITY Console software.

ENVIRONMENT

  • ACQUITY TUV
  • ACQUITY Console

PROCEDURE

  1. Replace the lamp as described in the ACQUITY UPLC TUV Overview and Maintenance Guide.

  2. Power-on the detector, and then wait about 30 minutes for the lamp to warm before resuming operations.

  3. Tip: Cycling power to the detector (that is, powering-off and then powering-on the instrument) initiates the verification procedures.

  4. Result: The lamp’s serial number and installation date are automatically entered into the Lamp Change Record table.

    Alternatively:  

  5. In the ACQUITY console, select Maintain > Change Lamp.

  6. Click New Lamp.

  7. Type the serial number for the new lamp, and then click OK.

  8. The counter automatically resets to 0 hours of lamp life. You cannot manually reset the counter to 0 hours

ADDITIONAL INFORMATION

 