Loud noise and vibrations coming from Alliance degasser pump - WKB54366

Article number: 54366

SYMPTOMS

  • Degasser pump makes loud noise and vibrates
  • Degasser pressure struggles to reach 0.9 psi upon initialization

ENVIRONMENT

  • Alliance HPLC System

CAUSE

Degasser outlet line is in the waste container, restricting the degasser pump

FIX or WORKAROUND

  1. Remove the degasser outlet line from the waste container.

ADDITIONAL INFORMATION