515 Pump leaks at inlet fitting to check valve - WKB5343

Article number: 5343

SYMPTOMS

  • Leak at inlet fitting to check valve housing (RHS)
  • Unable to stop by tightening

ENVIRONMENT

  • 515 HPLC Pump 
  • 1515 HPLC Pump
  • 1525 HPLC Pump
  • 1525 EF HPLC Pump
  • 1525 micro HPLC Pump  
     

CAUSE

Worn or damaged fittings

FIX or WORKAROUND

Replace the tube or tubes. Left-hand tube = 430001253. Right-hand tube = 430001256

ADDITIONAL INFORMATION

Original part number WAS207081 changed to 430001253 in December 2007.
Original part number was WAS207082 changed to 430001256 in December 2007.