CM-A Advanced Mode: eCords are only recorded in column position 1 - WKB5334
Article number: 5334
SYMPTOMS
-
Column 2 would not read any column history. It appears that no column data is being stored.
ENVIRONMENT
-
H-Class ACQUITY Column Manager - Active
-
I-Class ACQUITY Column Manager - Active
-
ACQUITY UPLC Systems with 2D Technology
CAUSE
- Advanced mode turns off eCord functionality. All injections are being recorded to a single column position in advanced mode of operation.
- eCords are only recorded in column position 1.
FIX or WORKAROUND
- Comprehensive eCord support in advanced mode is not possible to fully manage when considering custom configurations from a software perspective.
- There is limited support for eCord in the advanced mode.
ADDITIONAL INFORMATION
CRI-7703 exists as an enhancement request to improve eCord functionality with CM-A in Advanced mode