With CM-A and CM-Aux in Advanced mode, column shutdown in method shuts down column positions only in the CM-A - WKB5332
Article number: 5332
SYMPTOMS
-
Shutdown using the "Shutdown All Columns" check box shuts down column positions only in the CM-A; configured CM-Aux modules do not shut down
-
Changing the programmed column temperature to "off" does not shut down CM-Aux column positions
ENVIRONMENT
-
ACQUITY H-Class Column Manager - Active
-
ACQUITY I-Class Column Manager - Active
-
ACQUITY UPLC Systems with 2D Technology
CAUSE
- Advanced mode turns off column shutdown functionality when comparing column select mode.
- The CM-Aux is not supported in the advanced mode of operation.
FIX or WORKAROUND
- Change the programmed temperature of the column positions to room temperature.
- Ensure that temperature preferences are programmed so that method temperature is applied to all column positions.
When running the shutdown method with a temperature programmed at room temperature, this temperature is applied to the column positions in the CM-Aux.