No signal with ACQUITY FLR detector - WKB53225
Article number: 53225
SYMPTOMS
- No signal or no data
- FLR sat idle after annual PM
- New lamp
- No leaks detected
ENVIRONMENT
- 2475 Multi-Wavelength Fluorescence Detector
- ACQUITY UPLC FLR Detector
CAUSE
Detector is saturated.
FIX or WORKAROUND
- Run fresh HPLC-grade water through the flow cell for 15 minutes.
- Reboot the detector.
- Perform normalization. (See screen shot and links below.)
2475 operator's guide / manual
ADDITIONAL INFORMATION
id53225, 2475, 2475B, A-10FL, A-30FL, UPBINARY, UPFLRARC, UPFLRDET, UPPFLR