Static leak test fails with ACQUITY UPLC BSM - WKB52844

Article number: 52844

SYMPTOMS

  • Static leak test fails for the primary pump

ENVIRONMENT

  • ACQUITY UPLC BSM

CAUSE

The i2V primary check valve cartridge is leaking.

FIX or WORKAROUND

  1. Replace the i2V check valve cartridge (700003563, 2/pk).
  2. Repeat the Dynamic leak test 

ADDITIONAL INFORMATION

i2V Cartridge, Hexane/THF (700004139, 2/pk)

id52844, checkvalve, DLT, UPBINARY, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM