Static leak test fails with ACQUITY UPLC BSM - WKB52844
Article number: 52844
SYMPTOMS
- Static leak test fails for the primary pump
ENVIRONMENT
- ACQUITY UPLC BSM
CAUSE
The i2V primary check valve cartridge is leaking.
FIX or WORKAROUND
- Replace the i2V check valve cartridge (700003563, 2/pk).
- Repeat the Dynamic leak test
ADDITIONAL INFORMATION
i2V Cartridge, Hexane/THF (700004139, 2/pk)