Non-English Windows OS regional setting causes higher injection volume in ACQUITY UPLC Sample Manager - WKB52717
Article number: 52717
SYMPTOMS
- With ACQUITY Sample Manager on partial loop setting, when you select 5 µL, it injects 50 µL instead
ENVIRONMENT
- Analyst 1.6.2
- Driver Pack 4 SR1
- ACQUITY UPLC Sample Manager
CAUSE
Regional settings are using a comma as decimal indicator.
FIX or WORKAROUND
- Change the regional settings to English (United States). In Windows OS, navigate to Control Panel > Clock, Language, and Region.
- In the Formats tab, set format to 'English (United States)' in the drop down.
- Click the "Additional settings..." button.
- In the Numbers tab, set "Decimal symbol" to a decimal point.
- Change "List separator" to a comma.
- Apply > OK.
ADDITIONAL INFORMATION
id52717, UPBINARY