Article number: 51795

SYMPTOMS

  • Dynamic leak test fails with leak rate above acceptable limit
  • Unable to compress

ENVIRONMENT

  • ACQUITY UPLC Binary Solvent Manager (BSM)
  • nanoACQUITY UPLC
  • ACQUITY UPLC I-Class BSM
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC Isocratic Solvent Manager (ISM)
  • ACQUITY Premier Binary Solvent Manager (BSM)
  • ACQUITY Premier Quaternary Solvent Manager (QSM)
  • ACQUITY Arc
  • ACQUITY UPLC M-Class

CAUSE

Pump head seals failure. 

FIX or WORKAROUND

  1. Replace seals in the failing pump head
  2. Pressurize the pump seals for at least 15 minutes to a pressure close to the test pressure (>10 000 psi)
  3. Confirm Dynamic leak test is now passing

ADDITIONAL INFORMATION

If the ACQUITY UPLC Quaternary or Binary Solvent Manager seal wash housing (see part locator for finding the part number according to module type) shows signs of etching or rust might need to be replaced along with the pump head.

Part numbers for the pump head rebuild kits:

Accumulator rebuild kit:  700004746

Passive check valve Primary rebuild kit:  700004748

i2V Primary rebuild kit: 700004747

I-Class Accumulator rebuild kit: 700010584

I-Class Primary rebuild kit:  700010583

ACQUITY Premier BSM Primary rebuild kit: 700010581

ACQUITY Premier BSM Accumulator rebuild kit: 700010582

M-Class Primary rebuild kit: 700011473

M-Class Accumulator rebuild kit: 700011474

 

 