What is the part number for the flow restrictor used to increase system backpressure on the ACQUITY BSM or H-Class QSM? - WKB5084

Article number: 5084

ENVIRONMENT

  • ACQUITY UPLC Binary Solvent Manager
  • ACQUITY I-Classs Binary Solvent Manager
  • ACQUITY H-Class Quaternary Solvent Manager
  • ACQUITY H-Class Bio Quaternary Solvent Manager

ANSWER

205000547 for the KIT, WATERS UPLC FLOW RESTRICTOR

ADDITIONAL INFORMATION

  • The Waters Flow Restrictor is a 40um x 35cm capillary with PEEK-SIL coating (inert).
  • Install the restrictor between the solvent manager mixer and the sample manager injector.
  • Expect the increase in system pressure to be greater than 2000psi on average using the Waters Flow Restrictor.
  • Manufacturing label Part Number 430002180

 