What gradient mixer can be used to increase the dwell volume on the Alliance 2695? - WKB5080
Article number: 5080
ENVIRONMENT
- Alliance 2695
ANSWER
- The "ASSY, SINGLE STAGE MIXER" (WAT051519) adds approximately 1.12mL of delay volume to the system.
- Install the gradient mixer between the outlet of the gradient proportioning valve (GPV) and the inlet of the primary actuator drive.
- Optional mixers that can be used with Alliance 2695:
| Part number | Description | Volume | Identified by number screws |
| WAT051519 | ASSY, SINGLE STAGE MIXER | 1.12 mL | 10 |
| WAT055847 | GRADIENT MIXER SS ASSY | 300 µl | 5 |
| WAT031785 | Inert Gradient Mixer Assy | 300 µl | 5 |
| WAT088453 | Prep Mixer | 7.5 mL | 6 |
ADDITIONAL INFORMATION
Part number WAT051518 for the 600 GRADIENT MIXER is used specifically with the 600 Pump (600 Fluidic Handling Unit).
This can also be used on Prep LC system running two 515 pumps