Skip to main content
Waters

How to switch between HTCH and 30-cm column heater/cooler on the same sytem configuration - WKB5078

Article number: 5078

OBJECTIVE or GOAL

To switch between the ACQUITY UPLC High-Temperature Column Heater (HTCH) and ACQUITY UPLC 30-cm Column Heater/Cooler (30-cm CHC) when both are in the system configuration.

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • ACQUITY UPLC High-Temperature Column Heater (HTCH)
  • ACQUITY UPLC 30-cm Column Heater/Cooler (30-cm CHC)

PROCEDURE

  • Both column compartments are connected using the Y cable, (700004442 Y-Cable, HD D-Sub)
  • Can only use one column compartment at a time
  • To change from one to the other:
    • To control the temperature of the HTCH:
      1. Power-off the 30-cm Column Heater/Cooler.
      2. Power-off the Sample Manager.
      3. Power-on the Sample Manager.
    • To control the temperature of the 30-cm CHC:
      1. Power-on the 30-cm Column Heater/Cooler.
      2. Power-on the Sample Manager.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.