Skip to main content
Waters

Loose needle latch on ACQUITY H-Class Sample Manager - WKB50174

Article number: 50174

SYMPTOMS

  • Seal Force Sensor H/W Fault (0.0)
  • Needle/Seal Move Limit Exceeded (5.0)
  • Zp-axis home h/w fault
  • Needle is loose from the needle carriage and bent/damaged

ENVIRONMENT

  • ACQUITY H-Class Sample Manager - Flow Through Needle (SM-FTN)

CAUSE

The latch on the needle carriage is loose; it rotates easily in either direction.

clipboard_ea60a38a9ff3d81763f6edbefe1597f54.png

See example of typical picture from a customer when issue is present:

clipboard_ed3f25592b8631dd62ce4b34ab4416f97.png

FIX or WORKAROUND

  1. Tighten the screw fitting located in the center of the needle carriage needle latch.
  2. Return the needle to the correct position.
  3. Perform needle Z-axis calibration, characterize needle seal, and needle seal leak test as appropriate.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.