Shifting retention times on ACQUITY - WKB49914
Article number: 49914
SYMPTOMS
- Shifting/unstable retention times in chromatographic run
- During dynamic leak test, primary/accumulator pump A or B unable to compress
- Dynamic leak test fails with too high leak rate on primary/accumulator pump A or B
ENVIRONMENT
- ACQUITY UPLC I-Class BSM
- ACQUITY UPLC H-Class BSM
- ACQUITY UPLC BSM
- ACQUITY APC Isocratic Solvent Manager
- ACQUITY PREMIER BSM
CAUSE
Leaking pump
FIX or WORKAROUND
ADDITIONAL INFORMATION
Primary pump head or seal wash plate to be considered if replacement seals fail. (WKB20628)
Applies also to Accumulator dynamic leak test failure
