What is the best way to remove air from the solvent lines of a 1525 pump after running out of mobile phase? - WKB4938

Article number: 4938

ENVIRONMENT

  • 515 HPLC Pump
  • 1515 Isocratic HPLC Pump
  • 1525 Binary HPLC Pump
  • 1525 Micro Binary HPLC Pump

ANSWER

Manually prime using a priming syringe: Insert the priming syringe into the Luer fitting at the center of the draw-off valve handle. Turn the handle counterclockwise, about 1/2-turn, to open the valve, and pull back on the plunger to remove air bubbles. Once the air is removed, close the valve and start flow. 

Manual Prime.JPG

Flush the system with 100% methanol for a minimum of 30 minutes. Note: If methanol is not miscible with your mobile phase, use an intermediate mobile phase.