Skip to main content
Waters

Low sensitivity from ACQUITY UPLC Sample Manager - FTN - WKB49059

Article number: 49059

SYMPTOMS

  • Lower than expected response from injections

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle (FTN)
  • ACQUITY UPLC Sample Manager - Fixed Loop (FL)
  • XEVO
  • SYNAPT
  • MS

CAUSE

The purge line is dry, so the sample syringe has air in its barrel, causing poor sample draw.

Or there is a small air bubble in the syringe.

FIX or WORKAROUND

Prime the sample purge line and make sure the line has no air bubbles in it.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.