Low sensitivity from ACQUITY UPLC Sample Manager - FTN - WKB49059

Article number: 49059

SYMPTOMS

  • Lower than expected response from injections

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle (FTN)
  • ACQUITY UPLC Sample Manager - Fixed Loop (FL)
  • XEVO
  • SYNAPT
  • MS

CAUSE

The purge line is dry, so the sample syringe has air in its barrel, causing poor sample draw.

Or there is a small air bubble in the syringe.

FIX or WORKAROUND

Prime the sample purge line and make sure the line has no air bubbles in it.

ADDITIONAL INFORMATION

 