No peaks observed on Alliance 2795 - WKB48673
Article number: 48673
SYMPTOMS
- No peaks observed following injection of sample on a 2795
ENVIRONMENT
- Alliance 2795
CAUSE
Deck offset calibration is off.
FIX or WORKAROUND
- In the instrument console, select "Diag".
- Select "Change needle". (The needle moves to the back of the XYZ assembly.)
- Press "Ok". (The needle moves back to the injector assembly.)
- The needle deck offset calibrates automatically.
ADDITIONAL INFORMATION