2414 RID monitors but does not start collecting with an injection - WKB4865
Article number: 4865
SYMPTOMS
- 2414 displays a baseline while in Monitor
- 2414 does not start when an injection is made
- 2487 starts collecting when an injection is made
ENVIRONMENT
- 2414 Refractive Index (RI) Detector
- Breeze
- Empower
- Manual injector
- 1515
- 2487
- 2695 when mixing IEEE and Ethernet modules
CAUSE
The I/O connections from the 2414 to the manual injector are set up for an Event and not Inject Start.
FIX or WORKAROUND
Enable an Event in the Instrument Method in Breeze for Time 0.0 Event 1 Pulse.
ADDITIONAL INFORMATION
id4865, 1515, 2414, 2414B, 2487, A-10RI, BREEZESW, SUPBZ