2414 internal temperature can't be set to 55 ℃ via Empower due to recognized as 410 rather than W2414 - WKB48434
Article number: 48434
SYMPTOMS
- The internal temperature of a 2414 Refractive Index Detector controlled by Empower can not be set to 55 ℃ by the software.
- The software should be able to set the internal temperature to 55 ℃ according to 2414 instrument manual.
ENVIRONMENT
- 2414
- Empower 3
- Driver Pack 2016 Release 1 or later not installed.
CAUSE
Empower recognizes the 2414 as in 410 mode. In this mode, the lower and upper limit of internal temperature setting is 30 and 50 ℃, respectively. So the highest temperature that can be set is only 50 ℃.
FIX or WORKAROUND
- Download Driver Pack 2016 Release 1 or later.
- Update 2414 firmware to 2.10 or higher.
- Update 2414 drivers to 1.65 or higher.
- Empower recognizes the refractive index detector as W2414, thus there will be no more temperature control problems.