No signal with Alliance 2695 , e-SAT/IN, and a non-Waters detector using Empower - WKB4848
Article number: 4848
SYMPTOMS
- No signal or data when using the e-SAT/IN, 2695, and ECD
- New instrument method does not solve problem
ENVIRONMENT
- Empower 3
- Alliance 2695
- e-SAT/IN
- Amtek ECD detector
CAUSE
An inject start cable is missing between the 2695 and e-SAT/IN.
FIX or WORKAROUND
- Connect an event start cable (inject cable) between the 2695 (position 1,2 on connector B) and the e-SAT/IN (position marked "inject start").
- See the e-SAT/IN Module Installation Guide and the 2695 Separations Module Operator's Guide.