2475 Fluorescence Detector fails SQT system readiness check - WKB47474

Article number: 47474

SYMPTOMS

  • 2475 Fluorescence Detector fails system readiness check

ENVIRONMENT

  • ACQUITY UPLC FLR Detector
  • 2475 Multi-Wavelength Fluorescence Detector

CAUSE

Inconsistence in height and area
systemreadinesscheck.jpg

FIX or WORKAROUND

  1. Calibrate the photomultiplier tube (PMT). See the 2475 Multi-wavelength Fluorescence Detector Overview and Maintenance Guide (715004756).

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.