"Pump H/W over pressure" error on nano ACQUITY Binary Solvent Manager - WKB47479

Article number: 47479

SYMPTOMS

Pump H/W over pressure errors on starting pump flow.

ENVIRONMENT

nanoACQUITY UPLC

ACQUITY UPLC M-Class

CAUSE

Powering-off the pump with high system pressure causes incorrect pump head transducer calibration on startup.

FIX or WORKAROUND

Wait for system pressure read back to reach zero before cycling power on pump.

ADDITIONAL INFORMATION