Volume increase into the vials on Alliance 2695 - WKB4732

Article number: 4732

SYMPTOMS

  • Volume increase into the vials during drawing sample
  • Sample syringe not loose
  • Solvent on top of vials or carousel

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

V1 valve leaking

FIX or WORKAROUND

Rebuild V1 valve.

ADDITIONAL INFORMATION

HP Valve Rebuild Kit (WAT045424)

HP Motorized Valve (WAT045438)