Eluent leaks from needle on 2695 separations module - WKB4728
Article number: 4728
SYMPTOMS
- Eluent leaks from needle on 2695 module
ENVIRONMENT
- 2695 Separations Module
CAUSE
Bad needle seal in the seal pack assembly
FIX or WORKAROUND
- Rebuild seal pack.
- Seal Pack Rebuild Kit w/Needle (WAT271019)
- The Seal Pack Rebuild Kit includes all of the parts necessary to replace seals in the seal body, including the high-pressure seals. Replacement of these seals requires the HPMV Tool Kit (WAT045427). The Rebuild Kit includes the following: one Teflon washer, one needlewash filter, one upper spacer, one lower spacer, one alignment tube, one wash tube seal, one needle, two high-pressure seals, one compression screw, and one ferrule.