ACQUITY FLR detector error: Improper Flow Cell Contents Possible - WKB46994
Article number: 46994
SYMPTOMS
- Instrument does not initialize
- "Improper Flow Cell Contents Possible" error appears during startup
ENVIRONMENT
- ACQUITY UPLC FLR Detector
CAUSE
The wrong lamp is installed.
FIX or WORKAROUND
- Power-off the fluorescence detector.
- Replace the lamp.
- Power-on the detector.
- Allow the instrument to initialize.