Alliance 2695 Separations Module breaks vials - WKB4535
Article number: 4535
SYMPTOMS
- Vials break during an injection
- No errors observed
ENVIRONMENT
- 2690/2695/e2695 Alliance HPLC
- Alliance Separations Module
- Waters vials with low-volume inserts
CAUSE
The default needle depth offset is too low for low-volume inserts.
FIX or WORKAROUND
You can offset the depth of the needle in the vial to compensate for when you use low-volume inserts in the sample vials.
From the front panel of the 2695, select Diag > Service Utilities > Custom Offset and enter the value listed in Table B-4 of the 2695 Separations Module Operator's Guide.
ADDITIONAL INFORMATION
Table B-4 Separations Module Vials and Low-Volume Inserts
id4535, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB