The 2998 PDA reports that the shutter is closed and will not open - WKB3813
Article number: 3813
SYMPTOMS
- When resetting through the Empower Run Samples control panel, the 2998 PDA reports that the shutter is closed and will not open
ENVIRONMENT
-
2998 PDA
-
Empower 3
CAUSE
The computer and the instruments are not communicating properly.
FIX or WORKAROUND
- Power-off the instruments.
- Shut down the computer.
- Cycle power on the Ethernet switch.
- Turn on the computer and allow it to reach the normal desktop display.
- Turn on the instruments.