2695: FATAL ERROR--In VrtxTask main Assertion failed: (sm) file sepmeth.cxx, line 1184 - WKB3643

Article number: 3643

SYMPTOMS

  • Upon initialization screen displays: “initializing”, then Error: --FATAL ERROR--In VrtxTask main Assertion failed: (sm) file sepmeth.cxx, line 1184 
  • Cannot go to the main screen
  • Restart does not resolve error
  • Clear +8 does not resolve error

 

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

This is a low level firmware issue relating to programmable events in Methods screen

FIX or WORKAROUND

  1. Perform a Clear +0 from the front panel.
    • This erases all methods, logs, user and instrument configurations, and seal pack data.
    • It does not clear the system serial number.

ADDITIONAL INFORMATION

You need to do the following after Clear +0:
1) Perform the seal pack calibration.
2) Type the IEEE address (if applicable) in system configuration in Empower.
3) Recreate all 2695 methods.

id3643, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB