ACQUITY M-Class console high pressure limit is unstable when Bar is selected as the unit of measurement - WKB32685
Article number: 32685
SYMPTOMS
- When using Bar as the unit of measurement, if a user sets a flow rate (µL/min) on the µBSM and a high pressure limit in Bar (e.g., 600), starting the flow leads to immediate overpressure as the high pressure limit changes to a very low value (e.g., 50 bar)
- If the high pressure limit is set in psi or kPa, the µBSM functions normally
ENVIRONMENT
- ACQUITY M-Class
- MassLynx
CAUSE
Unknown software bug: Defect CRI-4133
FIX or WORKAROUND
None at this time. A workaround is to use psi or kPa as the unit of measure.
ADDITIONAL INFORMATION
