600F transducer is not zeroing - WKB3255

Article number: 3255

SYMPTOMS

  • Transducer is not zeroing when trying to make an adjustment.
  • Pressure does not go below 1000 psi when trying to zero with no flow and tubing disconnected from transducer.
  • System faults with "Over pressure".
  • Pump transducer does not zero with a reboot.

ENVIRONMENT

  • 510 HPLC Pump
  • 600 pump
  • 600 isocratic pump

CAUSE

Bad transducer.

FIX or WORKAROUND

  1. Replace the Transducer, WAT060328

 

 

ADDITIONAL INFORMATION

The pressure the transducer reads with no flow and no tubing attached could be different (not 1000 psi), but not 0 psi.

id3255, 600C, 600F, 600S, 616F