600F transducer is not zeroing - WKB3255
Article number: 3255
SYMPTOMS
- Transducer is not zeroing when trying to make an adjustment.
- Pressure does not go below 1000 psi when trying to zero with no flow and tubing disconnected from transducer.
- System faults with "Over pressure".
- Pump transducer does not zero with a reboot.
ENVIRONMENT
- 510 HPLC Pump
- 600 pump
- 600 isocratic pump
CAUSE
Bad transducer.
FIX or WORKAROUND
- Replace the Transducer, WAT060328
- 600_Transducer_replacement.pdf
ADDITIONAL INFORMATION
The pressure the transducer reads with no flow and no tubing attached could be different (not 1000 psi), but not 0 psi.
