Alliance iS frit regularly blocks up - WKB324787
Article number: 324787
SYMPTOMS
- Alliance iS frit regularly blocks up:
- After the GPV
- Before the primary check valve
- Using buffered solutions
- System clears blockage with aqueous flushing
ENVIRONMENT
- Alliance iS HPLC System
CAUSE
Buffer salts precipitate in the frit when going from 100% buffer to 100% organic solution.
FIX or WORKAROUND
Change priming order to transition from buffer to water, instead of going straight to organic.