Alliance iS frit regularly blocks up - WKB324787

Article number: 324787

SYMPTOMS

  • Alliance iS frit regularly blocks up:
    • After the GPV
    • Before the primary check valve
  • Using buffered solutions
  • System clears blockage with aqueous flushing

ENVIRONMENT

  • Alliance iS HPLC System

CAUSE

Buffer salts precipitate in the frit when going from 100% buffer to 100% organic solution.

FIX or WORKAROUND

Change priming order to transition from buffer to water, instead of going straight to organic.

ADDITIONAL INFORMATION

See the article How to change the priming order on an Alliance iS.