BSM accumulator or primary "unable to compress". Dynamic Leak Test failure after pump maintenance - WKB32496
Article number: 32496
SYMPTOMS
- BSM pump failing the Dynamic Leak Test with error: "Accumulator unable to compress" or "Primary unable to compress"
- Recent pump head maintenance
ENVIRONMENT
- ACQUITY UPLC BSM
- ACQUITY UPLC H-Class QSM
CAUSE
PEEK washer/gasket on the primary check valve displaced during fitting and squashed out of shape.
FIX or WORKAROUND
- Remove check valve with improperly seated washer.
- Replace PEEK washer:
- 700005218 for the WASHER GASKET, HIGH PRESS, PEEK I2V - (for intelligent check valve i2V)
- 700005221 for the WASHER, CHECK VALVE, PEEK (for accumulator check valve)
- Prime system.
- Run dynamic leak test.
ADDITIONAL INFORMATION
To prevent the PEEK washer from displacing itself during check valve installation, lightly wet the washer with a small amount of methanol when installing it on the check valve. The methanol will create adhesion between the washer and check valve, preventing it from moving.