Peak tailing in chromatogram - wkb31081
Article number: 31081
SYMPTOMS
- Significant tailing on all peaks
- Fresh mobile phase, wash solvents, column, and frits
ENVIRONMENT
- ACQUITY UPLC
CAUSE
Void volume caused by incorrect tubing installation (incorrect ferrule depth or damaged ferrule)
FIX or WORKAROUND
- Inspect all tubing and fitting connections from the pump to the detector.
- Replace tubing or fittings that have damaged or incorrectly set ferrules.