Variable injections on an ACQUITY system - WKB30047

Article number: 30047

SYMPTOMS

  • Repeat injections show variable response
  • No peaks present in some injections

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • Arc HPLC
  • ACQUITY Arc

CAUSE

Air in the system

FIX or WORKAROUND

  1. Ensure that sufficient wash solvent is present and prime the system—syringe, wash, and solvent lines.
  2. Ensure that all air bubbles are removed from the injection system and that the purge line filter is fully submerged in solvent.