Error with 2489 and PDA detector : Data buffer overflow - WKB29907
Article number: 29907
SYMPTOMS
- "Data buffer overflow" on the front panel
ENVIRONMENT
- 2489 UV/Vis Detector
- Firmware 1.41
- Empower 3
CAUSE
PCS 30743
FIX or WORKAROUND
- Use the "reset detector" function in the Run Samples control panel for the detector.
- Ensure that the rest of the system is also in a ready state.
- Reboot the computer and the 2489.
- Upgrade the firmware when possible.
ADDITIONAL INFORMATION
- This issue is corrected in the 1.65 firmware and later
- This defect also effects firmware versions previous to 1.65 for the ACQUITY PDA and TUV detectors
id29907, 2489, 2489B, 2489SFC, A-10UV, A-30PDA, A-30UV, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, UPAPDA, UPARPPDA, UPARPTUV, UPC2PDA, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA, UPPTUV, UPTUV, UPTUVARC, UPTUVARCB, UPTUV-E, UPTUVTC