"Transducer h/w range error" occurs on ACQUITY UPLC H-Class QSM - WKB298976

Article number: 298976

SYMPTOMS

  • "Transducer h/w range error" when priming system
  • Random pressure spike (fluctuation) 

ENVIRONMENT

  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)

CAUSE

Leaking and corroded fittings from pump head to transducer

FIX or WORKAROUND

  1. Replace the fittings to and from the transducers (430002316 and 430001120).
  2. Prime the system.

ADDITIONAL INFORMATION

 