"Pump Bad State" error on Alliance iS - WKB294050
Article number: 294050
SYMPTOMS
- "Pump Bad State" error appears on startup
- Bypassing inlet filters and re-initializing solves the issue
ENVIRONMENT
- Alliance iS HPLC System
CAUSE
Blocked inlet line filter
FIX or WORKAROUND
Replace the inlet filter (700002913).
ADDITIONAL INFORMATION
Blocked inlet filters result from poor solvent practices—most likely aqueous mobile phase that was left on the system too long, or immiscible solvents.
Remember: Filters do not block themselves; they are always blocked by introduced contaminants.