How to bake a drift tube on an ACQUITY UPLC ELS or 2424 ELS detector - WKB293211
Article number: 293211
OBJECTIVE or GOAL
Bake an ELSD drift tube.
ENVIRONMENT
- 2424 Evaporative Light Scattering Detector
- ACQUITY UPLC ELS Detector
PROCEDURE
- Remove Column and attach a union
- Flow 100% HPLC grade water through the system
- Set the drift tube temperature to 100 °C.
- Set gas flow to 50 psi.
- "Bake" for at least one hour.
- Set LC flow to zero.
- Turn off temperature and let the gas flow for at least 30 minutes.