Low peak response on ACQUITY UPLC FLR or Arc 2475 detector - WKB292230

Article number: 292230

SYMPTOMS

  • Low peak response on ACQUITY UPLC FLR detector
  • Low peak response on Arc 2475 detector

ENVIRONMENT

  • ACQUITY UPLC FLR detector
  • 2475 detector

CAUSE

The lamp energy is low.

FIX or WORKAROUND

Replace the lamp for ACQUITY UPLC FLR (part 201000193) or 2475 (part 201000131).

ADDITIONAL INFORMATION