What are the ACQUITY sample compartment temperature specifications? - WKB2916

Article number: 2916

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • ACQUITY H-Class SM-FTN
  • ACQUITY I-Class SM-FTN
  • ACQUITY I-Class SM-FL
  • ARC HPLC FTN-R

ANSWER

ARC FTN-R have is designed on the same back ground as H-Class and I-Class. So the specification is the same for FTN-R as for the H-Class and I-Class

The specifications are as follows:

Sample Compartment Specification.jpg

Definitions:

  • Settable Range - Programmable limits of the software
  • Tolerance Range - Acceptable thermal target at the set point
  • Ambient - Room temperature measured at the ambient temperature sensor located behind the right-hand side panel of the sample manager
  • De-rating - Calculated target range when ambient temperatures are higher than the SM is designed for

ADDITIONAL INFORMATION

To get a better understanding of this specification, examine the following table:
Sample CompartmentRange.jpg

*Set point = 4C

Enhancement Request submitted to allow setting asymmetric alarm band upper and lower limits [CRI-9735]