ACQUITY Arc fails DLT on primary - WKB288685

Article number: 288685

SYMPTOMS

  • DLT failure on primary pump head

ENVIRONMENT

  • ACQUITY Arc QSM

CAUSE

Leak in the tubing from the transducer to the accumulator check valve

FIX or WORKAROUND

Replace the tubing. 

ADDITIONAL INFORMATION

There are two part numbers for this tube in GPL:

  • 700010681 - Tube Assembly, Transducer to Accumulator
  • 700011670 - Tube Assembly, Transducer to ACC CV

 If unsure, order both.