How to perform the sample and wash syringe leak test on an ACQUITY UPLC Sample Manager - WKB288642

Article number: 288642

OBJECTIVE or GOAL

How to perform the sample and wash syringe leak test on an ACQUITY UPLC Sample Manager

ENVIRONMENT

  • ACQUITY UPLC Sample Manager

PROCEDURE

  1. For the syringe leak tests - go to the Acquity Console and select the Sample Manager from the System tree.
  2. Select Maintain > Sample syringe leak test or Wash syringe leak test.
  3. The appropriate "Syringe leak test" window appears - select Start to begin test.
  4. When completed, the results window will appears reporting a Pass / Fail Status.